«TEST SAMPLE EXTRACTOR FL-MIKRO-LAB-KOMPAKT WITH MEASURED VALUE EVALUATION LAPTOP/PC INDEX SEITE: 2 TECHNICAL DATA 2 1. TEST SAMPLE EXTRACTOR ...»

-1–

________________________________________________________________________

TEST SAMPLE EXTRACTOR

FL-MIKRO-LAB-KOMPAKT WITH

MEASURED VALUE EVALUATION

LAPTOP/PC

INDEX

SEITE: 2 TECHNICAL DATA

2 1. TEST SAMPLE EXTRACTOR FL-MIKRO-LAB-KOMPAKT

2 1.1 CONSTRUCTION 2 1.2 STARTING UP 2 1.2.1 SOFTWARE INSTALLATION 2 1.2.2 HARDWARE CONNECTION 3 1.3 HANDLING 3 1.3.1 CALIBRATION – MATERIAL ASSIGNMENT 3-5 1.3.1.1 CALIBRATION BY WEIGHT 6-7 1.3.1.2 VOLUMETRICALLY CALIBRATION 8-9 1.3.1.3 MATERIAL ASSIGNMENT 9-10 1.3.2 MEASURE OPERATION 10 1.4 CLEANING 10 1.5 ERROR DETECTION 10 1.5.1 INEXACT MEASURE RESULTS 11 1.5.2 CHECK OF MOISTURE PROBE 11 1.6 INTERFERENCE WITH ISM-FREQUENCES Probennehmer FL-Mikro-Lab-Kompakt mit LAPTOP / PC Franz Ludwig Gesellschaft für Mess- und Regeltechnik mbH * Budenheimer Straße 1 * 55124 Mainz-Gonsenheim * Telefon (0 61 31) 910 46-0 * Fax (0 61 31) 910 46-24 * info@fludwig.com

-2– ________________________________________________________________________

TECHNICAL DATA

FL-MIKRO-LAB-KOMPAKT:

H: 170 mm / ∅: 110 mm Dimensions Cover Highly resistant aluminum Weight 2,8 kg Temperature Range of the material to measure 0-80°C 180 cm3 Volume Connection cable USB A / Length: 3 m Measuring principle Microwave / 433 MHz

SYSTEM REQUIREMENTS OF EVALUATION-PC:

Operation system WINDOWS 2000 SP4 / XP SP2 / VISTA 32-bit / 7 Serial Interface 1x USB 2.0

1. TEST SAMPLE EXTRACTOR FL-MIKRO-LAB-KOMPAKT

1.1 Construction

1.2 Starting Up 1.2.1 Software Installation Proceed SETUP.EXE from the included software CD and follow the steps.

An Icon is created automatically on the desktop.

1.2.2 Hardware connection Connect the test sample extractor to your PC using the included connection cable.

–  –  –

________________________________________________________________________

1.3 Handling

Important information:

Before real moisture values can be displayed, a calibration has to be proceeded or a pre-calibrated calibration graph has to be operated. When measuring material with different ingredients and grain structure, you have to take several calibration graphs. You have to make sure the correct calibration graph is chosen in the evaluation PC when starting the measurement. Calibration curves without limit can be stored.

Generals:

Language Setting: Start the program click on ‚FILE‘ and choose ‚OPTIONS‘. Set your language. This setting is valid until it is manually changed through the user or if the software is re-installed.

1.3.1 Calibration – material assignment

Important information:

You can choose two possibilities of calibration:

1. Calibration by weight–OUR RECOMMENDATION (Chapter 1.3.1.1)

2. Volumetrically calibration (Chapter 1.3.1.2) 1.3.1.1 Calibration by weight • Prepare three material samples to fill the FL-MIKRO-LAB-KOMPAKT during the calibration.

One sample should be at the lower and the other on the upper moisture range which can be expected during your measuring operation. (in this example 6% and 12%).

The third sample has to be dried to 0%.

• Pay attention that FL-MIKRO-LAB-KOMPAKT and PC are connected during the whole processing.

• Choose ‚CALIBRATION‘ tab and click on ‚NEW‘:

• Enter the material’s indication and the first (lower) moisture value.

–  –  –

________________________________________________________________________

• Fill the FL-MIKRO-LAB-KOMPAKT with the dry material (0% moisture).

• Remove the material from FL-MIKRO-LAB-KOMPAKT and weigh it to determine the filling weight. (In this example 115 g).

• Now weigh the same quantity from the material with the lower moisture (in this example: 115g of 6%) and fill the FL-MIKRO-LAB-KOMPAKT.

–  –  –

________________________________________________________________________

• Now, close the FL-MIKRO-LAB-KOMPAKT and click on „START“ to take the first moisture value.

• After the first value is taken, the following picture appears. Please enter the second moisture value (in this example 12%) and re-fill the container with this material. Proceed the same as for the first moisture value (6%). At the end the calibrated calibration graph appears. When the screen shows ‚FINISH’, confirm this calibration graph with „OK“.

…… • After choosing ‚OK‘, the graph is stored and is graphically displayed.

Continue with “1.3.1.3 Material Assignment”.

–  –  –

________________________________________________________________________

1.3.1.2 Volumetrically calibration • Prepare two material samples to fill the FL-MIKRO-LAB-KOMPAKT during the calibration.

One sample should be at the lower and the other on the upper moisture range which can be expected during your measuring operation. (in this example 6% and 12%). Remark: The second value should not be 0%.





• Pay attention that FL-MIKRO-LAB-KOMPAKT and PC are connected during the whole processing.

–  –  –

• Click now on the ‚NEW‘ button in the calibration graph field and enter the calibration graph’s indication and the first moisture value of the calibration.

• Fill now the FL-MIKRO-LAB-KOMPAKT with the dryer material (in this example 6%).

–  –  –

________________________________________________________________________

• Click now on „START“ to take the first moisture value.

• After the first moisture value has been taken, the following picture appears. Enter now the second moisture value (in this example 12%) and re-fill the container with this material. Proceed the same as for the first moisture value (6%). At the end you will receive the calibrated calibration graph. When the screen shows ‚FINISH’, confirm this calibration graph with „OK“.

… … …

–  –  –

________________________________________________________________________

• Now the calibrated and the stored calibration graph is graphically displayed on the indication ‚BARLEY’.

1.3.1.3 Material Assignment • Choose tab „MATERIAL CLAIM“ then click button „NEW“.

• Enter now the product indication and assign the calibrated calibration graph. Then confirm with „OK“.

–  –  –

________________________________________________________________________

• The assignment is completed by now and is added to the listing.

1.3.2 Measuring Operation Start the program… The currently chosen product (‚Product name‘) and the assigned calibration graph is shown at the upper part of the displayed window.

–  –  –

________________________________________________________________________

• Fill the FL-MIKRO-LAB-KOMPAKT and click the button ‚START‘ to start the measurement. The measured moisture is displayed in blue and even recorded simultaneously.

• All taken moisture values are listed at tab ‚PROTOCOL‘..

To display a certain measured value course, mark the requested and recorded values and click at the button ‚SHOW MEASURE PROGRESS‘.

1.4 CLEANING Make sure no material remains are soiling the test sample extractor nor the probe´s measure surface after emptying the FL-MIKROLAB-COMPACT. Material remains can be removed with a brush or air pressure. Material stockings on the probe´s surface can be removed with common household cleaners.

1.5 ERROR DETECTION

1.5.1 Inexact measure results • Calibration has be done?

• Did you take several material specific calibration curves for the measuring of very different materials (compound, structure)?

• Did you assign the corresponding calibration curve to the product designation?

• Soiling and build-ups on the probe´s measure surface and the inner sample container are impossible?

• Has the FL-MIKRO-LAB-KOMPAKT been filled completely before measurement?

• Is the material sample to measure homogeneously mixed every time?

• Are the moisture values reliable evaluated with the calibration or the comparative measurement through reference system (ex. Drying oven)? You probably have to proceed control dryings.

–  –  –

________________________________________________________________________

1.5.2 Check of moisture probe At tab MATERIAL CLAIM you have the possibility to get the probe´s measure channel displayed when touching the button CHECK PROBE. The displaying of the probe´s measure channel allows the function checking of the microwave moisture probe.

The “measuring channel” with uncovered probe has to lay about 2500 (+/-300).

When you fill the FL-MIKRO-LAB-KOMPAKT up to 50% with water, the value should fall to 400 (+/-200).

Should the values of your probe lay out of this range, please contact one of our technicians for further steps. The “Quotient” is a linking up with “measuring channel” and indicates the noncalibrated measured value.

1.6 INTERFERENCE WITH ISM-FREQUENCES

Radio frequency devices use particular ISM-Frequencies (Industrial Scientific and Medical), which are international prescribed.

Our Microwave Moisture Probe, which counts as this type of device, is operated inside the frequency range from 433,05 to 434,79 MHz.

Radio frequency ranges are utilized in many commonly known devices. Example of these devices as follows:

Radio Alarm Systems - Vehicle Opener Mobile Transmitters - Garage door Opener Remote Controls - Wireless Movement Detectors On account to the huge number of ISM-Devices it can lead to mutual influences or interferences, if they are operated in direct environment. The radio signals might be received from others as the expected receiver. To avoid each kind of interferences, we recommend, that your other operating ISM-devices, which use the range from 433,05 – 434,79 MHz as well, not be used in immediate vicinity of our Microwave Moisture Probes resp. be laid out for other ISM-Frequencies. This goes especially for possible danger zones as for example cranes, roll-doors, buckle conveyor belts, e.g. In researches and tests, which had been done by us and outside institutes, have been found out, that the from us used frequency range from 433,05 to 434,79 MHz is the most advantageous to get reliable and accurate Moisture Measurement with our Microwave Moisture Probes.

Please feel free to contact us. We look forward to answering any questions you may have.

–  –  –

Franz Ludwig Gesellschaft für Mess- und Regeltechnik mbH * Budenheimer Straße 1 * 55124 Mainz-Gonsenheim * Telefon (0 61 31) 910 46-0 * Fax (0 61 31) 910 46-24 * info@fludwig.com





Similar works:

«Fakultät Informatik Institut für Softwareund Multimediatechnik, Professur Softwaretechnologie Technische Universität Dresden, 01062 Dresden Prof. Dr.rer.nat.habil. Uwe Aßmann Klausur Softwaretechnologie WS 2014/15 Name: Vorname: Immatrikulationsnummer: Aufgabe Maximale Erreichte Punktanzahl Punktanzahl Gesamt Hinweise:  In der Klausur ist als Hilfsmittel lediglich ein A4-Blatt, beidseitig beschrieben, zugelassen.  Die Klammerung der Aufgabenblätter darf nicht entfernt werden. ...»

«August 2015 Market Perspective | Technical Analysis Passing the deflation parcel The US stock market is now finely balanced between bulls and bears in one of the narrowest trading ranges seen in over 20 years. European stocks continue to consolidate after the 40% October-April rally, though they are somewhat at the mercy of EUR/USD price action, with the euro stabilising against the dollar after its big fall. The oil price remains a key variable, and is now very close to a major technical...»

«Endometrial (Uterine) Cancer What is endometrial cancer? Endometrial cancer starts when cells in the inner lining of the uterus (endometrium) begin to grow out of control. Cells in nearly any part of the body can become cancer, and can spread to other areas of the body. To learn more about how cancers start and spread, see What Is Cancer? The picture below shows where the uterus is located. About the uterus and endometrium The uterus is a hollow organ, which is normally about the size and shape...»

«Synthesis and characterization of 8 arm-Poly(ethylene glycol) Based hydrogels via Michael addition or click chemistry for Biomedical Applications Vorgelegt von M.Sc. Zhenfang Zhang aus Henan Von der Fakultät II Mathematik und Naturwissenschaften der Technischen Universität Berlin zur Erlangung des akademischen Grades Doktor der Ingenieurwissenschaften Dr.-Ing. Genehmigte Dissertation Promotionsausschuss: Vorsitzender: Prof. Dr. Nediljko Budisa Berichter/Gutachter: Prof. Dr. Ir. Marga C....»

«CONCERN WORLDWIDE Responding to Pakistan’s Internally Displaced (RAPID) Fund Phase II Frequently Asked Questions (FAQs) This question-and-answer fact sheet is intended to help the applicant organization understand all aspects of the RAPID Fund. If an organization is applying for funding under the RAPID Fund, please also read the Proposal Development Guidelines available on our website, which provides guidance for completing the application. Please visit Concern’s RAPID Fund website for full...»

«Carbon Nanotube Devices Dissertation zur Erlangung des akademischen Grades Doktoringenieur (Dr.-Ing.) vorgelegt von Dipl.-Ing. Robert Viktor Seidel geboren am 08.08.1975 in Potsdam Fakultät Maschinenwesen der Technischen Universität Dresden 1. Gutachter: Prof. Dr. rer. nat. habil. Wolfgang Pompe Technische Universität Dresden 2. Gutachter: Prof. Dr. rer. nat. habil. Bernd Büchner IFW Dresden, Technische Universität Dresden 3. Gutachter: Dr. rer. nat. habil. Ehrenfried Zschech AMD Saxony...»

«Automotive Consulting Solution Exit Control Framework Agenda 1. Kundennutzen 2. Funktionsbeschreibung 3. Abbildung im System 4. Technischer Steckbrief © SAP SE or an SAP affiliate company. All rights reserved. 2 Kundennutzen Lösung  Erprobte Lösung/Service der SAP Automotive Beratung  Bereits bei mehreren Kunden im produktiven Einsatz  Lösung und Dokumentation sind in den Sprachversionen Deutsch und Englisch verfügbar Kosten  Exakte Kostenschätzung möglich. Einführung zum...»

«Data Migration: Moving from Dell PowerVault™ MD3000i/MD3000 to MD3200i/MD3220i and MD3600i/MD3620i Series Storage Arrays A Dell Technical White Paper PowerVault™ MD3200/MD3200i and MD3600i Storage Arrays Data Migration from the Dell PowerVault™ MD3000i/MD3000 to MD3200i/MD3220i and MD3600i/MD3620i Series Storage Arrays THIS WHITE PAPER IS FOR INFORMATIONAL PURPOSES ONLY, AND MAY CONTAIN TYPOGRAPHICAL ERRORS AND TECHNICAL INACCURACIES. THE CONTENT IS PROVIDED AS IS, WITHOUT EXPRESS OR...»

«Product Brief Storage Virtualization isn’t About Storage Date: January 2010 Author: Steve O’Donnell, Senior Analyst Data Center Practice Abstract: Virtualization is at the center of all 21st Century IT systems, yet many CIOs fail to fully understand all of the benefits it can deliver to the data center operation. When we think of virtualization, we think compute, network, and storage—and we mostly think about driving up utilization on each. Storage controllers have always offered the...»

«Reaktionstechnische Untersuchungen zur Ethylenherstellung aus wässrigen Ethanollösungen (Fermenterausträgen) Eine nachhaltige Alternative zum Steam-Cracker Abschlussbericht für die Max-Buchner-Forschungsstiftung Januar 2011 Kennziffer 2783 Gökhan Aras (1. Förderperiode) Heiner Busch (2. Förderperiode) Prof. Dr.-Ing. Herbert Vogel Ernst-Berl-Institut für Technische und Makromolekulare Chemie Technische Universität Darmstadt Petersenstraße 20, 64287 Darmstadt 1. Abstract Die...»





 
<<  HOME   |    CONTACTS
2016 www.book.dislib.info - Free e-library - Books, dissertations, abstract

Materials of this site are available for review, all rights belong to their respective owners.
If you do not agree with the fact that your material is placed on this site, please, email us, we will within 1-2 business days delete him.